From: Different Methods for Measurements and Estimation of Pulse Wave Velocity are not Interchangeable
Measurement pairs PWV (m/s) | ΔM* | p |
---|---|---|
SphygmoCor—Arteriograph 24 h | − 0.307 | 0.091† |
SphygmoCor—Mobil-O-Graph 24 h | 1.670 | 0.000† |
SphygmoCor- ePWV | 1.870 | 0.089† |
Arteriograph 24 h—Mobil-O-Graph 24 h | 1.947 | 0.000† |