- Conference Abstract
- Open access
- Published:
P60 Invasive Validation of Antares, a New Algorithm to Calculate Central Blood Pressure from Oscillometric Upper Arm Pulse Waves
Artery Research volume 25, page S101 (2019)
Abstract
Background
Antares is an algorithm for pulse wave analysis (PWA) by oscillometric blood pressure monitors in order to estimate central (aortic) blood pressure (cBP). Antares aims to enable brachial cuff-based blood pressure (BP) monitors to be type II-devices, determining absolute cBP values independently of potential peripheral BP inaccuracies. The present study is an invasive validation of the Antares algorithm.
Methods
We followed entirely the 2017 ARTERY protocol for validation of non-invasive cBP devices, the 2013 ANSI/AAMI/ ISO 81060-2 and 2018 AAMI/ESH/ISO validation standard protocols. In total, 191 patients undergoing cardiac catheterization were included, of which 145 patients entered analysis. Invasive cBP recordings were prospectively compared to simultaneous non-invasive cBP estimations using Antares, integrated into an oscillometric BP monitor. Antares analyses pulse waves during deflation of the cuff (patent of Redwave Medical GmbH). Generally speaking, pulse waves of a normal oscillometric BP measurement are taken for PWA with no need for altering standard BP pump operation.
Results
Mean difference between invasive and non-invasively estimated systolic cBP was 0.71 mmHg with standard deviation of 5.95 mmHg, fulfilling the highest validation criteria. Bland-Altman plot reveals good limits of agreement.
Conclusion
Antares is the first algorithm for estimation of cBP that entirely fulfills the 2017 ARTERY and AAMI/ESH/ISO validation protocols including criteria for high accuracy devices. The Antares algorithm turns the oscillometric upper arm blood pressure monitor into a type II-device for estimation of true cBP. Integration of Antares into BP monitors could enable the measurement of PWA parameters in virtually every practice in future.

Bland altman of estimated vs. invasive central SBP
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).
About this article
Cite this article
Doerr, M., Richter, S., Eckert, S. et al. P60 Invasive Validation of Antares, a New Algorithm to Calculate Central Blood Pressure from Oscillometric Upper Arm Pulse Waves. Artery Res 25 (Suppl 1), S101 (2019). https://doiorg.publicaciones.saludcastillayleon.es/10.2991/artres.k.191224.091
Published:
Issue Date:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.2991/artres.k.191224.091