P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms
Artery Research volume 25, page S162 (2019)
Abstract
Background
Calibration of carotid artery tonometry waveforms needs to account for pulse pressure augmentation [1]. Due to its location nearer to the heart, carotid pulse pressure (carPP) is about 10 mmHg lower than brachial (brachPP) [1]. In practice, perfect applanation is hard to achieve with tonometry and thus tracking-errors may occur, leading to an exaggerated fall in the diastolic limb of the tonometry waveform. Consequently, the mean-to-diastolic difference of the waveform is underestimated, leading to an overestimation of carPP after calibration. We assessed the relationship between brachPP and carPP in 100 subjects, aged 18–80 yrs.
Methods and Results
Brachial pressure (Omron) and carotid tonometry (Sphygmocor) measurements were performed by a single, highly skilled vascular technician, according to Artery Society methodological standards. CarPP was 5 ± 13 mmHg lower than brachPP. Bland-Altman analysis revealed a positive trend (r = 0.695, p < 0.0001) between the carPP–brachPP difference and their mean, with carPP being 20 mmHg lower than brachPP at a mean 40 mmHg which extended to carPP being 20 mmHg higher(!) than brachPP at a mean 80 mmHg. Accordingly, 30% of the study population showed unphysiological carPP, indicating an unrealistic i.e. reverse amplification. Even with a tolerance of +5 mmHg, still 20% of cases showed unrealistic values.
Conclusion
In practice, calibrated carotid tonometry waveforms may yield unphysiological overestimations of carotid pulse pressure. This finding implies (potential) estimation errors in regression model studies on central pressure and local stiffness, suggesting brachial pulse pressure would be a valid compromise.
References
Kelly R, Fitchett D. Noninvasive determination of aortic input impedance and external left ventricular power output: a validation and repeatability study of a new technique. J Am Coll Cardiol 1992;20:952–63.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).
About this article
Cite this article
Reesink, K., Marais, L. & Boutouyrie, P. P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms. Artery Res 25 (Suppl 1), S162 (2019). https://doiorg.publicaciones.saludcastillayleon.es/10.2991/artres.k-191224.148
Published:
Issue Date:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.2991/artres.k-191224.148