P.41 Measurement of Pressure-dependent Intra-Beat Changes in Carotid Pulse Wave Velocity using Image-Free Fast Ultrasound
Artery Research volume 26, page S63 (2020)
Abstract
Background
The clinical significance of pressure-dependent intra-beat changes in local pulse-wave velocity (C) has recently come to light [1]. While reported methods require arterial pressure and diameter measurements from a single site to assess intra-beat changes in C, we present an image-free fast ultrasound device that performs this by capturing diameter waveforms from two proximal locations on an artery.
Methods
The functionality was assessed on eight normotensive participants (26 ± 4 years). By perturbing blood pressure through a short duration moderate lower body negative pressure intervention [2], CD and CF pulse wave velocities corresponding to diastolic and 80% of peak pressure were measured from the carotid artery. Human NIBP system (ADInstruments, India) was used for monitoring continuous pressure.
Results
The device captured dual-diameter waveforms and evaluated CD and CF, with a beat-to-beat variation <8% during baseline. CD was smaller than CF (p < 0.001), 4.2 ± 0.5 m/s versus 4.6 ± 0.7 m/s during baseline and 3.7 ± 0.6 m/s versus 3.9 ± 0.7 m/s during intervention. Concomitant to the drop in group-average diastolic (17%) and systolic (18%) pressures during the intervention, CD and CF dropped by 14% and 16%, respectively. The statistically significant correlation (r > 0.6, p < 0.001) of CD and CF with the diastolic and systolic pressures for each individual was preserved even after adjusting for heart-rate.
Conclusions
The device demonstrated its functionality and reliably measured the incremental nature of C. Its pressure-dependent intra-beat variations and inter-beat dynamics during the intervention concurred with literature. Further studies are underway to demonstrate the potential use of the device in vascular research and clinical applications.
References
Nabeel PM, Raj VK, Joseph J, Abhidev VV, Sivaprakasam M. Local pulse wave velocity: theory, methods, advancements, and clinical applications. IEEE Rev Biomed Eng 2020;13:74–112.
Goswami N, Blaber AP, Hinghofer-Szalkay H, Convertino VA. Lower body negative pressure: physiological effects, applications, and implementation. Physiol Rev 2019;99:807–51.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).
About this article
Cite this article
Raj, K.V., Nabeel, P.M., Joseph, J. et al. P.41 Measurement of Pressure-dependent Intra-Beat Changes in Carotid Pulse Wave Velocity using Image-Free Fast Ultrasound. Artery Res 26 (Suppl 1), S63 (2020). https://doiorg.publicaciones.saludcastillayleon.es/10.2991/artres.k.201209.053
Published:
Issue Date:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.2991/artres.k.201209.053